# 2025-10787: FDA Streamlines a Test That Checks Many Antibiotics at Once

2025-10787, Medical Devices; Immunology and Microbiology Devices; Classification of the Cellular Analysis System for Multiplexed Antimicrobial Susceptibility Testing, is a Federal Register rule from Health and Human Services Department, Food and Drug Administration published 2025-06-13, effective 2025-06-13; PRIA reads it as touching household money (as of 2026-10-30).

- Canonical: https://policyrisk.com/federal-register/2025-10787
- As of: 2026-10-30
- Plain-text twin of the page above; the HTML and this file carry the same facts.

## Facts
- Document number: 2025-10787
- Title: Medical Devices; Immunology and Microbiology Devices; Classification of the Cellular Analysis System for Multiplexed Antimicrobial Susceptibility Testing
- Type: Rule
- Agencies: Health and Human Services Department; Food and Drug Administration
- Citation: 90 FR 24962
- Published: 2025-06-13
- Effective: 2025-06-13
- Comments close: not stated
- Significant: no
- Touches household money: yes
- Regulation identifier numbers: none
- Last updated: 2026-10-30

## In one line

FDA Streamlines a Test That Checks Many Antibiotics at Once

## Summary

The FDA is officially putting the cellular analysis system for multiplexed antimicrobial susceptibility testing into Class II, meaning it now has special safety rules but fewer hurdles than the strictest category. This change helps companies bring these devices to patients faster and safer, starting June 13, 2025. Labs and medical device makers will benefit from clearer guidelines and smoother approval processes without extra costs.

## Abstract

The Food and Drug Administration (FDA, the Agency, or we) is classifying the cellular analysis system for multiplexed antimicrobial susceptibility testing into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the cellular analysis system for multiplexed antimicrobial susceptibility testing's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

## Effects PRIA read

- Device moved to Class II (easier access): benefit, medium. You may get faster access to this type of antimicrobial susceptibility test because FDA classified the cellular analysis system for multiplexed antimicrobial susceptibility testing as Class II (special controls), effective June 13, 2025 (classification applicable February 23, 2017). FDA states this…
- De Novo gives predicate status, easing future approvals: benefit, medium. If you make these devices, FDA's De Novo classification means this device can serve as a predicate for future devices, so other sponsors can use the less-burdensome 510(k) process instead of submitting another De Novo request or a premarket approval application. FDA explains this predicate status f…
- Manufacturers must meet specific special controls: cost, medium. If you make or market this device, you must comply with the listed special controls to be Class II and avoid automatic Class III placement. The order requires detailed design verification/validation (device description, primer/probe details, analytical studies including limit of detection, inclusiv…

## Related documents

- [2026-20848](https://policyrisk.com/federal-register/2026-20848): Use of Salt Substitutes To Reduce the Sodium Content in Standardized Foods, 2026-10-13
- [2026-16296](https://policyrisk.com/federal-register/2026-16296): Substances Generally Recognized as Safe, 2026-08-11
- [2026-14073](https://policyrisk.com/federal-register/2026-14073): Drug Establishment Registration and Drug Listing Requirements for Establishments Engaged in Distributed Manufacturing and Certain Foreign E…, 2026-07-13
- [2026-13047](https://policyrisk.com/federal-register/2026-13047): Establishment Registration and Product Listing for Tobacco Products, 2026-06-29
- [2025-04978](https://policyrisk.com/federal-register/2025-04978): Nonprescription Drug Product With an Additional Condition for Nonprescription Use, 2025-03-25
- [2025-01840](https://policyrisk.com/federal-register/2025-01840): Nonprescription Drug Product With an Additional Condition for Nonprescription Use, 2025-01-27

## Sources
- [2025-10787 on federalregister.gov](https://www.federalregister.gov/documents/2025/06/13/2025-10787/medical-devices-immunology-and-microbiology-devices-classification-of-the-cellular-analysis-system)
- [2025-10787 as published (PDF)](https://www.govinfo.gov/content/pkg/FR-2025-06-13/pdf/2025-10787.pdf)

## Related
- [Previous: 2025-10786, Medical Devices; Immunology and Microbiology Devices; Classification of the Dev…](https://policyrisk.com/federal-register/2025-10786)
- [Next: 2025-10788, Medical Devices; Immunology and Microbiology Devices; Classification of the Cli…](https://policyrisk.com/federal-register/2025-10788)
- [All Federal Register documents](https://policyrisk.com/federal-register)