# 2026-20725: Blood Cell Kit Finally Gets Official Safety Stamp

2026-20725, Medical Devices; Immunology and Microbiology Devices; Classification of the Hematopoietic Cell Enrichment Kit, is a Federal Register rule from Health and Human Services Department, Food and Drug Administration published 2026-10-09, effective 2026-10-09; PRIA reads it as touching household money (as of 2026-10-16).

- Canonical: https://policyrisk.com/federal-register/2026-20725
- As of: 2026-10-16
- Plain-text twin of the page above; the HTML and this file carry the same facts.

## Facts
- Document number: 2026-20725
- Title: Medical Devices; Immunology and Microbiology Devices; Classification of the Hematopoietic Cell Enrichment Kit
- Type: Rule
- Agencies: Health and Human Services Department; Food and Drug Administration
- Citation: 91 FR 64609
- Published: 2026-10-09
- Effective: 2026-10-09
- Comments close: not stated
- Significant: no
- Touches household money: yes
- Regulation identifier numbers: none
- Last updated: 2026-10-16

## In one line

Blood Cell Kit Finally Gets Official Safety Stamp

## Summary

The FDA is officially putting the hematopoietic cell enrichment kit into Class II, meaning it now has special safety rules but fewer hurdles than the highest-risk devices. This change helps make sure the kit is safe and effective while speeding up patient access to this cool medical tech. The new classification took effect on October 9, 2026, but has been in place since November 6, 2023, so manufacturers and patients can benefit now without extra costs.

## Abstract

The Food and Drug Administration (FDA) is classifying the hematopoietic cell enrichment kit into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the hematopoietic cell enrichment kit. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

## Effects PRIA read

- Device moved to Class II (easier access): benefit, medium. The FDA officially classified the hematopoietic cell enrichment kit as Class II (special controls). The order is effective October 9, 2026, and the classification has been applicable since November 6, 2023; FDA says this reduces regulatory burdens and will enhance patient access to the device.
- Device subject to 510(k) premarket notification: cost, medium. FDA states that hematopoietic cell enrichment kits are subject to premarket notification under section 510(k). Manufacturers must therefore prepare and submit 510(k) filings rather than being exempt from that requirement.
- Sample-collection devices must be FDA-cleared or included: cost, low. Any sample collection device used with the kit must be FDA-cleared, approved, or classified as 510(k) exempt for the sample types intended, or the sample collection device must be cleared as part of the kit's premarket submission.
- Minimum performance: 95% accurate call rate requirement: cost, medium. The intended-use statement must list specimen types for each malignancy and document acceptable enrichment using the minimum volume required to ensure a 95 percent accurate call rate when the cell concentration is at the device limit of detection, per the device labeling.
- Labeling must include specific limiting statements (48‑hour rule): mixed, low. Labeling must include detailed procedures and prominent limiting statements, including that bone marrow stability is severely compromised after 48 hours (or another FDA‑supported timeframe), that the device is only for enriching specimens for further analysis, that results do not provide informatio…
- Extensive verification and validation study requirements: cost, high. Manufacturers must provide detailed design verification and validation documentation, including recovery studies, a limit-of-detection study using an FDA-acceptable flow cytometry reference, reproducibility studies with multiple operators and three reagent lots across low/medium/high target levels,…
- De Novo class can serve as predicate for future devices: benefit, medium. FDA notes the De Novo classification allows the device type to be used as a predicate for future devices, enabling other sponsors to use the less burdensome 510(k) process rather than a De Novo or premarket approval application.

## Related documents

- [2026-20848](https://policyrisk.com/federal-register/2026-20848): Use of Salt Substitutes To Reduce the Sodium Content in Standardized Foods, 2026-10-13
- [2026-16296](https://policyrisk.com/federal-register/2026-16296): Substances Generally Recognized as Safe, 2026-08-11
- [2026-14073](https://policyrisk.com/federal-register/2026-14073): Drug Establishment Registration and Drug Listing Requirements for Establishments Engaged in Distributed Manufacturing and Certain Foreign E…, 2026-07-13
- [2026-13047](https://policyrisk.com/federal-register/2026-13047): Establishment Registration and Product Listing for Tobacco Products, 2026-06-29
- [2025-04978](https://policyrisk.com/federal-register/2025-04978): Nonprescription Drug Product With an Additional Condition for Nonprescription Use, 2025-03-25
- [2025-01840](https://policyrisk.com/federal-register/2025-01840): Nonprescription Drug Product With an Additional Condition for Nonprescription Use, 2025-01-27

## Sources
- [2026-20725 on federalregister.gov](https://www.federalregister.gov/documents/2026/10/09/2026-20725/medical-devices-immunology-and-microbiology-devices-classification-of-the-hematopoietic-cell)
- [2026-20725 as published (PDF)](https://www.govinfo.gov/content/pkg/FR-2026-10-09/pdf/2026-20725.pdf)

## Related
- [Previous: 2026-20722, Airworthiness Directives; Airbus Helicopters Deutschland GmbH (AHD) Helicopters](https://policyrisk.com/federal-register/2026-20722)
- [Next: 2026-20726, Medical Devices; Cardiovascular Devices; Classification of the Infant Pulse Rat…](https://policyrisk.com/federal-register/2026-20726)
- [All Federal Register documents](https://policyrisk.com/federal-register)