R.459.566 Fluoroscopic X-ray systems: Limitation of useful beam; spot filming.
2.5K chars
NAC 459.566 Fluoroscopic X-ray systems: Limitation of useful beam; spot filming. (NRS 459.201) All fluoroscopic X-ray systems must meet the following requirements: 1. The fluoroscopic tube must not produce X-rays unless the primary protective barrier is in position to intercept t…
R.459.568 Fluoroscopic X-ray systems: Activation of the tube.
0.5K chars
NAC 459.568 Fluoroscopic X-ray systems: Activation of the tube. (NRS 459.201) X-ray production in the fluoroscopic mode must be controlled by a device which requires continuous pressure by the fluoroscopist for the entire time of any exposure. When recording serial fluoroscopic i…
R.459.570 Fluoroscopic X-ray systems: Exposure rate limits; exception.
4.9K chars
NAC 459.570 Fluoroscopic X-ray systems: Exposure rate limits; exception. (NRS 459.030, 459.201) 1. The exposure measured at the point where the center of the useful beam enters the patient must not exceed 10 roentgens (100 millisieverts) per minute, except during recording of flu…