R 333.5371 Purpose and scope. Rule 371. (1) This part establishes requirements governing the use of x-radiation in dentistry. (2) This part applies to all registrants who use x-radiation in dentistry for the intentional exposure of humans. (3) In addition to the requirements of t…
R.5372 Scope
0.2K chars
R 333.5372 Scope. Rule 372. R 333.5373 to R 333.5376 apply to installations consisting of a single x-ray source, its individual control unit, and protective enclosure used for the production of intraoral radiographs. History: 2016 AACS.
R.5373 X-ray equipment
4.4K chars
R 333.5373 X-ray equipment. Rule 373. (1) The tube housing shall be of the diagnostic type. (2) The aluminum equivalent of the total filtration in the useful beam shall not be less than that shown in table 373-1. TABLE 373-1 Operating Potential Minimum Total Filter (Inherent plus…